The results of this pooled analysis show an association of IVUS guidance during percutaneous coronary intervention with better outcomes in patients with LM disease undergoing revascularization with DES.
Background:
The use of poly-
l
-lactide acid-based bioresorbable scaffolds is limited in daily clinical practice because of safety concerns and lack of physiological benefit. Magnesium-based bioresorbable scaffold (MgBRS) presents a short resorption period (<1 year) and have the potential of being thromboresistant and exhibiting early restoration of vasomotor function. To date, however, no randomized clinical trial has investigated the performance of MgBRS. Therefore, this study aimed to compare the in-stent/scaffold vasomotion between MgBRS and permanent metallic sirolimus-eluting stent (SES) at 12-month follow-up in ST-segment–elevation myocardial infarction patients.
Methods:
This investigator-driven, multicenter, randomized, single-blind, controlled trial randomized ST-segment–elevation myocardial infarction patients 1:1 to SES or MgBRS at 11 academic centers. The primary end point was the rate of increase (≥3%) after nitroglycerin in mean lumen diameter of the in-stent/scaffold segment at 12 months with superiority of MgBRS over SES in the as-treated population. The main secondary end points included angiographic parameters of restenosis, device-oriented composite end point, their individual components, and device thrombosis rate. Besides, endothelial-dependent vasomotor response to acetylcholine (ie, endothelial function) was also assessed in a subgroup of patients (n=69).
Results:
Between June 2017 and June 2018, 150 ST-segment–elevation myocardial infarction patients were randomized (MgBRS, n=74; SES, n=76). At 1 year, the primary end point was significantly higher in the MgBRS arm (56.5% versus 33.8%;
P
=0.010). Conversely, late lumen loss was significantly lower in the SES group (in-segment: 0.39±0.49mm versus 0.02±0.27mm,
P
<0.001; in-device: 0.61±0.55mm versus 0.06±0.21mm;
P
<0.001). The device-oriented composite end point was higher in the MgBRS arm driven by an increase in ischemia-driven target lesion revascularization rate (12[16.2%] versus 4[5.2%],
P
=0.030). Definite thrombosis rate was similar between groups (1[1.4%] in the MgBRS arm versus 2[2.6%] in the SES group;
P
=1.0). Endothelial function assessment at device segment evidenced a more pronounced vasoconstrictive response to maximal dose of acetylcholine in the MgBRS arm (−8.3±3.5% versus −2.4±1.3% in the SES group,
P
=0.003).
Conclusions:
When compared to SES, MgBRS demonstrated a higher capacity of vasomotor response to pharmacological agents (either endothelium-independent or endothelium-dependent) at 1 year. However, MgBRS was associated with a lower angiographic efficacy, a higher rate of target lesion revascularization, without thrombotic safety concerns.
Clinical Trial Registration:
URL:
https://www.clinicaltrials.gov
. Unique identifier: NCT03234348.
Bioresorbable vascular scaffolds have better conformability than conventional MPS. The clinical significance of the observed differences will require further investigation.
The hemodynamic microenvironment appears to regulate neointimal response following an Absorb BVS implantation. These findings underline the role of the ESS patterns on vessel wall healing and should be taken into consideration in the design of bioresorbable devices.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.