Different oxygen tensions do not significantly affect glucose measured with the GD-based amperometric test strips, and have minimal effect on GO-based photometric test strips. Increases in oxygen tension lowered glucose measured with GO-based amperometric test strips. We recommend that the effects of different oxygen tensions in blood samples on glucose measurements be minimized by using oxygen-independent test strips for point-of-care glucose testing in critically ill and other patients with high or unpredictable blood Po2 levels.
The performance of GD-based, oxygen-insensitive, handheld glucose testing was technically suitable for arterial specimens in critical care patients, cord blood and heelstick specimens in neonates, and capillary and venous specimens in other patients. Multicenter findings benchmark the performance of bedside glucose testing devices. With the new +/-15 mg/dL --> 100 mg/dL --> +/-15% accuracy criterion, point-of-care systems for handheld glucose testing should score 95% (or better), as compared with the recommended reference method. Physiologic changes, preanalytical factors, confounding variables, and treatment goals must be taken into consideration when interpreting glucose results, especially in critically ill patients, for whom arterial blood glucose measurements will reflect systemic glucose levels.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.