Micro-computed tomography (µCT) provides quantitative three-dimensional information of bone around titanium implants similar to classical histology. The study, based on an animal model, using cuboid-shaped biofunctionalised Ti6Al4V implants with surrounding bone after 4 weeks, is performed using 3 µCT-systems with X-ray tubes, one synchrotron-radiation-based µCT-system (SRµCT), and classical histology. Although the spatial resolution of the µCT-systems is comparable, only the results of SRµCT agree with results of classical histology. The X-ray tube sources give rise to huge artefacts in the tomograms (interface scattering, beam hardening), which impaired the quantitative analysis of bone up to about 200 µm from the implant surface. Due to the non-destructive character of µCT the specimens can be subsequently examined by classical histology without restriction. The quantitative comparison of bone formation uncovers the strong dependence of the detected amount of newly formed bone from the selected slice. This implies the necessity of 3D analysis. SRµCT and classical histology prove that surface modifications of the titanium implant significantly influence the bone formation. Using SRµCT, the preparation artefacts due to cutting and polishing are excluded.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.