Under the auspices of the Inter-American Metrology System (SIM), the National Institute of Standards and Technology (NIST) initiated a regional comparison for type K thermocouples from (100 to 1,100) • C with 11 participating countries. The use of type K material above approximately 200 • C is considered destructive. Therefore, each participating laboratory was sent new, unused wire from a lot of material characterized by NIST. The uniformity of the lot was remarkable, especially at temperatures above 500 • C; the standard deviation of the thermocouple emf values of multiple cuts tested at NIST was 2.7 µV or less over the full temperature range. The high uniformity eliminated any need to correct for variations of the transfer standard among the laboratories, greatly simplifying the analysis. The level of agreement among the laboratories' results was quite good. Even though test procedures and equipment varied significantly among the participants, the standard deviation of all emf values at each test temperature was less than the equivalent of 0.20 • C at 200 • C and below, and less than 0.60 • C from (400 to 1,100) • C. Of the 380 total bilateral combinations of the
Type K thermocouples are one of the most commonly used temperature sensors in industry. The skills, personnel, and facilities necessary for calibrating type K thermocouples are also applicable to the calibration of other base metal thermocouples, and, to a lesser extent, calibration of platinum–rhodium alloy thermocouples. Under the auspices of the Inter-american Metrology System (SIM), the National Institute of Standards and Technology (NIST) initiated a regional comparison for type K thermocouples from 100 °C to 1100 °C, with 11 participating countries. The use of type K material above approximately 200 °C is considered destructive. Therefore, each participating laboratory was sent new, unused wire from a lot of material characterized by NIST. The uniformity of the lot was remarkable, especially at temperatures above 500 °C; the standard deviation of the thermocouple emf values of multiple cuts tested at NIST was 2.7 µV or less over the full temperature range. The high uniformity eliminated any need to correct for variations of the transfer standard among the laboratories, greatly simplifying the analysis. The level of agreement among the laboratories' results was quite good. Of the 380 total bilateral combinations of the data at the eight test temperatures, only 13 (i.e., 3.4% of all combinations) are outside the k = 2 limits, and of these 13, only 3 are outside k = 3 limits. All of the outliers occur at temperatures of 800 °C and below, which suggests that drift of the type K wire due to high-temperature oxidation did not cause changes in thermocouple emf comparable to or larger than the claimed uncertainties.Main text.
To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the SIM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.