A fast and sensitive method for the determination of atropine, an alkaloid closely related to cocaine, is proposed. The principles of on-line ion-pair formation of alkaloid-metal complexes and liquid-liquid extraction are applied to the chemiluminescence determination of atropine. On mixing with a reversed micellar medium of cetyltrimethylammonium chloride in dichloromethane-cyclohexane (1:1 v/v)-water (0.3 M Na2CO3) containing luminol, the ion-pair complex of tetrachloroaurate(III) with atropinium produced an analytical chemiluminescence signal when it entered the reversed micellar water pool. Using the reverse-flow injection and chemical conditions optimized for atropine in aqueous samples, a detection limit of 1 ng/mL was achieved and a linear calibration graph was obtained with a wide dynamic range from 10 ng/mL to 100 micrograms/mL. The proposed method is simple and provides a good precision with a relative standard deviation (n = 6) of ca. 3% at the atropine concentration of 100 ng/mL. After a preliminary study involving the potential interference from species of organic, inorganic, and metallic nature, the method was applied to the determination of atropine in artificial urine samples and of atropine and scopolamine in pharmaceutical formulations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.