This publication details CMOS foundry fabrication, reliability stress assessment, and packaged sensor test results obtained during qualification of the SensL B-Series silicon photomultiplier (SiPM). SiPM sensors with active-area dimensions of 1, 3, and 6 mm were fabricated and tested to provide a comprehensive review of SiPM performance highlighted by fast output rise times of 300 ps and photon detection efficiency of greater than 41%, combined with low afterpulsing and crosstalk. Measurements important for medical imaging positron emission tomography systems that rely on time-of-flight detectors were completed. Results with LSYO:Ce scintillation crystals of 3 × 3 × 20 mm 3 demonstrated a 225 AE 2-ps coincidence resolving time (CRT), and the fast output is shown to allow for simultaneous acquisition of CRT and energy resolution. The wafer level test results from ∼150 k 3-mm SiPM are shown to demonstrate a mean breakdown voltage value of 24.69 V with a standard deviation of 0.073 V. The SiPM output optical uniformity is shown to be AE10% at a single supply voltage of 29.5 V. Finally, reliability stress assessment to Joint Electron Device Engineering Council (JEDEC) industry standards is detailed and shown to have been completed with all SiPM passing. This is the first qualification and reliability stress assessment program run to industry standards that has been reported on SiPM.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.