The CAS neonatal NIRS system determines absolute regional brain tissue oxygen saturation (SnO2) and brain true venous oxygen saturation (SnvO2) non-invasively. Since NIRS-interrogated tissue contains both arterial and venous blood from arterioles, venules, and capillaries, SnO2 is a mixed oxygen saturation parameter, having values between arterial oxygen saturation (SaO2) and cerebral venous oxygen saturation (SvO2). To determine a reference for SnO2, the relative contribution of SvO2 to SaO2 drawn from a brain venous site vs. systemic SaO2 is approximately 70:30 (SvO2:SaO2). If the relationship of the relative average contribution of SvO2 and SaO2 is known and does not change to a large degree, then NIRS true venous oxygen saturation, SnvO2, can be determined non-invasively using SnO2 along with SaO2 from a pulse oximeter.
Objective
A state-of-the-art centrifugal pump combined with hollow-fiber oxygenator for extracorporeal membrane oxygenation has potential advantages such as smaller priming volumes and decreased potential to cause tubing rupture as compared with the traditional roller head/silicone membrane systems. Adoption of these state-of-the-art systems has been slow in neonates as a result of past evidence of severe hemolysis that may lead to renal failure and increased mortality. Extracorporeal systems have also been linked to platelet dysfunction, a contributing factor toward intracranial hemorrhage, a leading cause of infant morbidity. Little data exist comparing the centrifugal systems with the roller systems in terms of hemolysis and platelet aggregation at low flow rates commonly used in neonatal extracorporeal membrane oxygenation.
Design
Prospective, comparative laboratory study.
Setting
University research laboratory.
Subjects
Centrifugal pump, roller pump, hollow-fiber oxygenator, and silicone membrane oxygenator.
Interventions
Comparative study using two pumps, the centrifugal Jostra Rotaflow (Maquet, Wayne, NJ) and the roller-head (Jostra, Maquet, Wayne, NJ), and two oxygenators, polymethly-pentene Quadrox-D (Maquet) and silicone membrane (Medtronic, Minneapolis, MN). Five test runs of four circuit combinations were examined for hemolysis and platelet aggregation during 6 hrs of continuous use in a simulated in vitro extracorporeal membrane oxygenation circuit circulating whole swine blood at 300 mL/min.
Measurements and Main Results
Hemolysis was assessed by spectrophometric measurement of plasma-free hemoglobin. Platelet aggregation was evaluated using monoclonal CD61 antibody fluorescent flow cytometry profiles. All of the extracorporeal membrane oxygenation systems created plasma-free hemoglobin at a similar rate compared with static blood control. There was no difference in the mean normalized index of hemolysis of the centrifugal/hollow-fiber oxygenator system as compared with the roller-head/silicone membrane systems (0.0032 g/100 L vs. 0.0058 g/100 L, p ≥ .7). None of the extracorporeal membrane oxygenation systems had a significant increase in platelet aggregation above baseline.
Conclusions
In a low-flow neonatal environment, a state-of-the-art centrifugal pump combined with new fiber-type oxygenators appear to be safe in regard to hemolysis and platelet aggregation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.