Biochemical analyses were performed on osteoarthritic and periprosthetic synovial fluid in order to propose changes to lubricant specifications currently outlined in orthopaedic wear testing standards. Osteoarthritic and periprosthetic synovial fluid samples were obtained from the hip and knee joints of 40 patients. The samples in each group were analysed and compared in order to identify differences between the protein concentration, constituent fractions, osmolality, thermal stability and the hyaluronic acid concentration and molecular weight distribution of osteoarthritic and periprosthetic synovial fluid. The average total protein concentration was approximately 30 g/L, which was much higher than the 20 g/L currently specified in the knee wear testing standard; however, the 30 g/L protein concentration matched the recently revised standard for hip simulator wear testing. No significant difference was found between the protein concentration, osmolality, thermal stability, and hyaluronic acid concentration of osteoarthritic and periprosthetic synovial fluid. The clinical data provided should be used to better define the composition of a more clinically relevant lubricant for orthopaedic wear testing.
Lubricant protein concentration is known to affect crosslinked polyethylene wear in in vitro testing; however, the biochemical nature of these lubricants may also have a significant effect on wear and dictate its clinical relevance. A modified approach to pin-on-disc testing was implemented to explore the effect of four biochemically different lubricants on the wear of two types of crosslinked polyethylene materials (XLK™ and Marathon™; DePuy Synthes, Warsaw, IN, USA). XLK was associated with higher wear rates than Marathon. In comparison to lubricants containing deionized water, lubricants containing phosphate buffered saline solution and hyaluronic acid increased osmolality by up to 1.2 times and thermal stability by up to 1.4 times. This biochemical change reduced wear by up to 12.5 times. Wear rates for XLK and Marathon differed by a factor of 3.2 using lubricants with phosphate buffered saline solution as the dilutive media, but only 2.0 for lubricants with deionized water. Interestingly, varying the concentration of hyaluronic acid did not have a significant effect on wear, and differences between XLK and Marathon wear rates were not found to be statistically significant when hyaluronic acid was added to the lubricant. The findings of this study showed that increasing the osmolality and thermal stability of lubricants to more clinical levels decreased wear; however, the effect of hyaluronic acid on wear may not be apparent in simplistic pin-on-disc testing. It was suggested that phosphate buffered saline solution be used as the dilutive media of choice in order to better differentiate the ranking of materials while maintaining some clinical relevance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.