Initiated by complaints of image artifacts, a thorough visual and radiographic investigation of 197 Fuji, 35 Agfa, and 37 Kodak computed radiography (CR) cassettes with imaging plates (IPs) in clinical use at four radiology departments was performed. The investigation revealed that the physical deterioration of the cassettes and IPs was more extensive than previously believed. It appeared that many of the image artifacts were the direct result of premature wear of the cassettes and imaging plates. The results indicate that a quality control program for CR cassettes and IPs is essential and should include not only cleaning of the cassettes and imaging plates on a regular basis, but also visual and radiographic image inspection to limit the occurrence of image artifacts and to prolong the life cycle of the CR equipment.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.