The energy response function of the semi-imaging ISA Riber MAC 2 electron spectrometer has been determinated over the kinetic energy range 150-1500 eV. The absolute transmission of the analyser is proportional to KE;", with n ranging from 1.02 to 0.43 and K from 1575 to 23.2 wben the focusing voltage VFl of the retarding field grid increases from 5 to 20 V, i.e. wben the analyser resolution increases from 055 to 2.2 eV. The power factor n and the proportionality factor K are linearly decreasing when the focusing voltage is increasing: n = 1.23-0.043 VF, and K = 211-108 VF,.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.