A venous cuff at the venous anastomosis of PTFE arteriovenous grafts for hemodialysis reduced the incidence of thrombotic occlusions; stenosis at the venous anastomosis was reduced. However, this did not result in a better patency rate. Therefore, the venous cuff should not be used routinely. Initial vein diameter and local problems (edema, obesity, or skin atrophy) appear to be the most important risk factors for graft failure.
The use of PTFE grafts and other types of vascular accesses has increased at the expense of BC fistulas. Recirculation and venous pressure measurements are primarily done when problems occur and not according to a standard surveillance schedule. For visualization of failing fistulas, 79% of the centres uses Duplex ultrasound analysis and 100% angiography. The popularity of PTA has increased from 46 to 92% of the centres.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.