We investigated the clinical usefulness of a newly developed flat-panel detector (FPD) system by comparing its physical imaging properties and low-contrast detectability with those of a current FPD system. The newly developed CsI-based indirect FPD (Canon, CXDI-40C) and current Gd(2)O(2)S-based FPD (Canon CXDI-11) systems were used. Characteristic curves, resolution properties, radiographic noise, detective quantum efficiencies (DQEs) and low-contrast detectability for both systems were measured. The new FPD system showed considerably lower noise levels than those of the current FPD system. DQE (0) s of the new and current FPD systems were 75% and 35%, respectively. Observer performance tests of the contrast-detail (C-D) phantom indicated that the new FPD system can significantly improve low-contrast performance over that obtainable with the current FPD system under the same conditions of exposure. The new FPD system provided approximately 50% reduction in exposure while providing comparable detectability. The newly developed FPD system provides radiographic images with excellent inherent physical image quality and low-contrast performance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.