A preconcentration module operated in flow mode and integrated with a sequential injection system with spectrophotometric detection was developed. Using the system, preconcentration was performed in continuous mode and was based on a membraneless evaporation process under diminished pressure. The parameters of the proposed system were optimized and the system was tested on the example of the spectrophotometric determination of Cr(III). The preconcentration effectiveness was determined using the signal enhancement factor. In the optimized conditions for Cr(III), it was possible to obtain the signal enhancement factors of around 10 (SD: 0.9, n = 4) and determine Cr(III) with precision and intermediate precision of 8.4 and 5.1% (CV), respectively. Depending on the initial sample volume, signal enhancement factor values of about 20 were achieved. Applicability of the developed preconcentration system was verified in combination with the capillary electrophoresis method with spectrophotometric detection on the example of determination of Zn in certified reference materials of drinking water and wastewater. Taking into account the enhancement factor of 10, a detection limit of 0.025 mg L−1 was obtained for Zn determination. Zn was determined with precision less than 6% (CV) and the results were consistent with the certified values.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.