The combination of functional MRI (fMRI) and MRS is a promising approach to relate BOLD imaging to neuronal metabolism, especially at high field strength. However, typical scan times for GABA edited spectroscopy are of the order of 6‐30 min, which is long compared with functional changes observed with fMRI.The aim of this study is to reduce scan time and increase GABA sensitivity for edited spectroscopy in the human visual cortex, by enlarging the volume of activated tissue in the primary visual cortex. A dedicated setup at 7 T for combined fMRI and GABA MRS is developed. This setup consists of a half volume multi‐transmit coil with a large screen for visual cortex activation, two high density receive arrays and an optimized single‐voxel MEGA‐sLASER sequence with macromolecular suppression for signal acquisition.The coil setup performance as well as the GABA measurement speed, SNR, and stability were evaluated. A 2.2‐fold gain of the average SNR for GABA detection was obtained, as compared with a conventional 7 T setup. This was achieved by increasing the viewing angle of the participant with respect to the visual stimulus, thereby activating almost the entire primary visual cortex, allowing larger spectroscopy measurement volumes and resulting in an improved GABA SNR. Fewer than 16 signal averages, lasting 1 min 23 s in total, were needed for the GABA fit method to become stable, as demonstrated in three participants. The stability of the measurement setup was sufficient to detect GABA with an accuracy of 5%, as determined with a GABA phantom. In vivo, larger variations in GABA concentration are found: 14‐25%. Overall, the results bring functional GABA detections at a temporal resolution closer to the physiological time scale of BOLD cortex activation.
To bring metabolic imaging based on multi-NMR toward practical use from the RF hardware perspective. Methods: A highly integrated RF coil is designed for whole-brain MRI and MRS targeted to five nuclear species: 1 H, 19 F, 31 P, 23 Na, and 13 C. Dipole antennas and closely loaded local receiver loops are combined in this setup. Results: High-quality in vivo scan results of 1 H, 31 P, 23 Na, and 13 C on healthy volunteers have been achieved. For 1 H, the transmit efficiency is 77% of a single-tuned commercial head coil (NOVA 8-transmit [Tx]/32-receive [Rx]; NOVA Medical, Wilmington, MA, USA). For 31 P, 110% SNR of a dual-tuned close-fit head-birdcage was achieved at the center of the subject, based on MR experiments on a phantom. For 31 P, 23 Na, and 13 C, bench measurements indicate SNR loss of 15%, 27%, and 30% compared with single-tuned conditions. 19 F performance has been proven to be similar to that of 1 H through bench tests and electromagnetic simulations. Conclusion: With this device, 1 H-based anatomic images that are expected to meet clinical requirements, as well as high-quality multi-NMR images and spectra, can be acquired within one scan session without hardware replacement or patient repositioning, enabling morphologic and metabolic MRI within acceptable scan time.
Immobilization masks are used to prevent patient movement during head and neck (H&N) radiotherapy. Motion restriction is beneficial both during treatment, as well as in the pre-treatment simulation phase, where MRI is often used for target definition. However, the shape and size of the immobilization masks hinder the use of regular, close-fitting MRI receive arrays. In this work, we developed a mask-compatible 8-channel H&N array that consists of a single-channel baseplate on which the mask can be secured and a flexible 7-channel anterior element that follows the shape of the mask. The latter uses high impedance coils to achieve its flexibility and radiolucency. A fully-functional prototype was manufactured, its radiolucency was characterized, and the gain in imaging performance with respect to current clinical setups was quantified. Dosimetry measurements showed an overall dose change of -0.3%. Small, local deviations were up to -2.7% but had no clinically significant impact on a full treatment plan, as gamma pass rates (3 %/3 mm) only slightly reduced from 97.9% to 97.6% (clinical acceptance criterion: ≥95%). The proposed H&N array improved the imaging performance with respect to three clinical setups. The H&N array more than doubled (+123%) and tripled (+246%) the signal-to-noise ratio with respect to the clinical MRI-simulation and MR-linac setups, respectively. G-factors were also lower with the proposed H&N array. The improved imaging performance resulted in a clearly visible SNR improvement of clinically used TSE and DWI acquisitions. In conclusion, the 8-channel H&N array improves the imaging performance of MRI-simulation and MR-linac acquisitions, while dosimetry suggests that no clinically significant dose changes are induced.
Quantitative three‐dimensional (3D) imaging of phosphorus (31P) metabolites is potentially a promising technique with which to assess the progression of liver disease and monitor therapy response. However, 31P magnetic resonance spectroscopy has a low sensitivity and commonly used 31P surface coils do not provide full coverage of the liver. This study aimed to overcome these limitations by using a 31P whole‐body transmit coil in combination with a 16‐channel 31P receive array at 7 T. Using this setup, we determined the repeatability of whole‐liver 31P magnetic resonance spectroscopic imaging (31P MRSI) in healthy subjects and assessed the effects of principal component analysis (PCA)‐based denoising on the repeatability parameters. In addition, spatial variations of 31P metabolites within the liver were analyzed. 3D 31P MRSI data of the liver were acquired with a nominal voxel size of 20 mm isotropic in 10 healthy volunteers twice on the same day. Data were reconstructed without denoising, and with PCA‐based denoising before or after channel combination. From the test–retest data, repeatability parameters for metabolite level quantification were determined for 12 31P metabolite signals. On average, 31P MR spectra from 100 ± 25 voxels in the liver were analyzed. Only voxels with contamination from skeletal muscle or the gall bladder were excluded and no voxels were discarded based on (low) signal‐to‐noise ratio (SNR). Repeatability for most quantified 31P metabolite levels in the liver was good to excellent, with an intrasubject variability below 10%. PCA‐based denoising increased the SNR ~ 3‐fold, but did not improve the repeatability for mean liver 31P metabolite quantification with the fitting constraints used. Significant spatial heterogeneity of various 31P metabolite levels within the liver was observed, with marked differences for the phosphomonoester and phosphodiester metabolites between the left and right lobe. In conclusion, using a 31P whole‐body transmit coil in combination with a 16‐channel 31P receive array at 7 T allowed 31P MRSI acquisitions with full liver coverage and good to excellent repeatability.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.