Though the 4 mm Gamma Knife helmet is used routinely, there is disagreement in the Gamma Knife users community on the value of the 4 mm helmet relative output factor. A range of relative output factors is used, and this variation may impair observations of dose response and optimization of prescribed dose. To study this variation, measurements were performed using the following radiation detectors: silicon diode, diamond detector, radiographic film, radiochromic film, and TLD cubes. To facilitate positioning of the silicon diode and diamond detector, a three-dimensional translation micrometer was used to iteratively determine the position of maximum detector response. Positioning of the films and TLDs was accomplished by manufacturing custom holders for each technique. Results from all five measurement techniques indicate that the 4 mm helmet relative output factor is 0.868 +/- 0.014. Within the experimental uncertainties, this value is in good agreement with results obtained by other investigators using diverse techniques.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.