Sevoflurane provided quicker emergence and early recovery compared with halothane, but the incidence of delirium was greater in preschool boys after sevoflurane.
These results indicate that Group I and II drugs exert some antioxidant activity in vitro, as measured by their protection of fluorescence decay of B-PE. Careful consideration of these properties might, then, serve to facilitate more efficient drug application.
The boys who were anxious before anesthesia showed a significantly greater incidence of problematic behavior on emergence from halothane anesthesia, compared with the boys who were calm before anesthesia.
Background: The STG‐22™ is the only continuous blood glucose monitoring system currently available. The aim of this study is to determine the accuracy and reliability of the STG‐22™ for continuously monitoring blood glucose level in post‐surgical patients.
Methods: Fifty patients scheduled for routine surgery were studied in surgical intensive care unit (ICU) of a university hospital. After admission to the ICU, the STG‐22™ was connected to the patients. An attending physician obtained blood samples from a radial arterial catheter. Blood glucose level was measured using the ABL™800FLEX immediately after blood collection at 0, 4, 8, and 16 h post‐admission to the ICU (total of 200 blood glucose values).
Results: The correlation coefficient (R2) was 0.96. In the Clarke error grid, 100% of the paired measurements were in the clinically acceptable zone A and B. The Bland and Altman analysis showed that bias±limits of agreement (percent error) were 0.04(0.7)±0.35(6.3) mmol (mg/dl) (7%), −0.11(−2)±1.22(22) (15%) and −0.33(−6)±1.28(23) (10%) in hypoglycemia (<70(3.89) mmol (mg/dl), normoglycemia (3.89(70)–10(180) mmol (mg/dl), and hyperglycemia (>10(180) mmol (mg/dl), respectively.
Conclusions: The STG‐22™ can be used for measuring blood glucose level continuously and measurement results are consistent with intermittent measurement (percentage error within 15%). Therefore, the STG‐22™ is a useful device for monitoring in blood glucose level in the ICU for 16 h.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.