A number of general conclusions are evident: (1) small field OR(det) are very sensitive to the simulated source parameters, and therefore, rigorous Monte Carlo linac model commissioning, with respect to measurement, must be pursued prior to use, (2) backscattered dose to the monitor chamber should be included in simulated OR(det) calculations, (3) the corrections required for diode detectors are design dependent and therefore detailed detector modeling is required, and (4) the reported detector specific correction factors may be applied to experimental small field OR(det) consistent with those presented here.
Commercial multileaf collimator (MLC) systems can employ leaves with rounded ends. Treatment planning beam modelling should consider the effects of transmission through rounded leaf ends to provide accurate dosimetry for IMRT treatments delivered with segmented MLC. We determined that an MLC leaf gap reduction of 1.4 mm is required to obtain an agreement between calculated and measured profile 50% dose points. A head and neck dosimetry phantom, supplied by the Radiological Physics Center (RPC), was planned and irradiated as a necessary credentialing requirement for the RTOG H-0022 protocol. The agreement between the RPC TLD measurements and treatment planning calculations was within experimental error for the primary and secondary planning target volumes (PTVs); however, the calculated mean dose for the critical structure was approximately 9% lower than the RPC TLD measurements. RPC radiochromic film profile measurements also indicated significant discrepancies (>5%) with calculated values especially in the high dose gradient region in the vicinity of the critical structure. These results substantiate our own in-house phantom measurements, performed with the same IMRT fields as for the RPC phantom experiment, using Kodak EDR2 film to measure absolute dose. Our results indicate a maximum underestimate of calculated dose of 12% with no leaf gap reduction. The discrepancy between measured and calculated phantom values is reduced to +/- 5% when a leaf gap reduction of 1.4 mm is used. A further improvement in the accuracy of dose calculation is not possible without a more accurate modelling of the leaf end transmission by the planning system. In the absence of published dosimetric criteria for IMRT our results stress the need for stringent in-house dosimetric QA and validation for IMRT treatments. We found the dosimetric validation service provided by the RPC to be a valuable component of our IMRT validation efforts.
The goal of this work was to examine the use of simplified diode detector models within a recently proposed Monte Carlo (MC) based small field dosimetry formalism and to investigate the influence of electron source parameterization has on MC calculated correction factors. BEAMnrc was used to model Varian 6 MV jaw-collimated square field sizes down to 0.5 cm. The IBA stereotactic field diode (SFD), PTW T60016 (shielded) and PTW T60017 (un-shielded) diodes were modelled in DOSRZnrc and isocentric output ratios (OR(fclin)(detMC)) calculated at depths of d = 1.5, 5.0 and 10.0 cm. Simplified detector models were then tested by evaluating the percent difference in (OR(fclin)(detMC)) between the simplified and complete detector models. The influence of active volume dimension on simulated output ratio and response factor was also investigated. The sensitivity of each MC calculated replacement correction factor (k(fclin,fmsr)(Qclin,Qmsr)), as a function of electron FWHM between 0.100 and 0.150 cm and energy between 5.5 and 6.5 MeV, was investigated for the same set of small field sizes using the simplified detector models. The SFD diode can be approximated simply as a silicon chip in water, the T60016 shielded diode can be modelled as a chip in water plus the entire shielding geometry and the T60017 unshielded diode as a chip in water plus the filter plate located upstream. The detector-specific (k(fclin,fmsr)(Qclin,Qmsr)), required to correct measured output ratios using the SFD, T60016 and T60017 diode detectors are insensitive to incident electron energy between 5.5 and 6.5 MeV and spot size variation between FWHM = 0.100 and 0.150 cm. Three general conclusions come out of this work: (1) detector models can be simplified to produce OR(fclin)(detMC) to within 1.0% of those calculated using the complete geometry, where typically not only the silicon chip, but also any high density components close to the chip, such as scattering plates or shielding material is necessary to be included in the model, (2) diode detectors of smaller active radius require less of a correction and (3) (k(fclin,fmsr)(Qclin,Qmsr)) is insensitive to the incident the electron energy and spot size variations investigated. Therefore, simplified detector models can be used with acceptable accuracy within the recently proposed small field dosimetry formalism.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.