The current paper aims at assessing the sensitivity of muscle and intervertebral disc force computations against potential errors in modeling muscle attachment sites. We perturbed each attachment location in a complete and coherent musculoskeletal model of the human spine and quantified the changes in muscle and disc forces during standing upright, flexion, lateral bending, and axial rotation of the trunk. Although the majority of the muscles caused minor changes (less than 5%) in the disc forces, certain muscle groups, for example, quadratus lumborum, altered the shear and compressive forces as high as 353% and 17%, respectively. Furthermore, percent changes were higher in the shear forces than in the compressive forces. Our analyses identified certain muscles in the rib cage (intercostales interni and intercostales externi) and lumbar spine (quadratus lumborum and longissimus thoracis) as being more influential for computing muscle and disc forces. Furthermore, the disc forces at the L4/L5 joint were the most sensitive against muscle attachment sites, followed by T6/T7 and T12/L1 joints. Presented findings suggest that modeling muscle attachment sites based on solely anatomical illustrations might lead to erroneous evaluation of internal forces and promote using anatomical datasets where these locations were accurately measured. When developing a personalized model of the spine, certain care should also be paid especially for the muscles indicated in this work.
Purpose
Tether breakage was reported as the most common complication of vertebral body tethering. However, as the literature suggests the physiological loads do not have the potential to cause the failure of the tether. Currently, the biomechanical reason behind the tether breakage is unknown. The current study aims to elucidate the effects of the tension forces on the failure mechanisms of the VBT and provide mechanical justification for how it can be identified radiographically.
Methods
Tensile tests (20%/min strain rate) were performed on single-unit VBT samples. Failure modes and mechanical characteristics were reported.
Results
The failure took place prematurely due to the slippage of the tether at the screw–tether junction where the tether is damaged significantly by the locking cap. Slippage was initiated at 10–13% tensile strain level where the tensile stress and tension force were 50.4 ± 1.5 MPa and 582.2 ± 30.8 N, respectively.
Conclusion
The failure occurs because of high-stress concentrations generated within the locking region which damages the tether surface and leads to the slippage of the tether. We observed that the loads leading to failure are within the physiological limits and may indicate the high likelihood of the tether breakage. The failure mode observed in our study is shown to be the dominant failure mode, and a design improvement on the gripping mechanism is suggested to avoid failure at the screw–tether junction. We observed that the tether elongates 10–13% prior to the breakage, which can be employed as a diagnostic criterion to screen for tether breakages radiographically.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.