In this paper, we report the design, fabrication, and testing of a lab-on-a-chip based microfluidic device for application of trapping and measuring the dielectric properties of microtumors over time using electrical impedance spectroscopy (EIS). Microelectromechanical system (MEMS) techniques were used to embed opposing electrodes onto the top and bottom surfaces of a microfluidic channel fabricated using Pyrex substrate, chrome gold, SU-8, and polydimethylsiloxane. Differing concentrations of cell culture medium, differing sized polystyrene beads, and MCF-7 microtumor spheroids were used to validate the designs ability to detect background conductivity changes and dielectric particle diameter changes between electrodes. The observed changes in cell medium concentrations demonstrated a linear relation to extracted solution resistance (Rs), while polystyrene beads and multicell spheroids induced changes in magnitude consistent with diameter increase. This design permits optical correlation between electrical measurements and EIS spectra.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.