Background: We compared the accuracy and practicability of a new combined ear sensor device measuring pulse oximetry and transcutaneous carbon dioxide tension. Methods: Validation studies were done by comparing the results of the combined sensor with arterial blood gas measurements. In an observational part, monitoring data were obtained from 25 patients undergoing colonoscopy, sedated with midazolam and alfentanil and from 8 patients without sedation. Results: There was an excellent correlation between the oxygen saturation and carbon dioxide tension measurements comparing the combined sensor with arterial blood gas analysis (R 0.96 and 0.93, respectively). A mean rise in transcutaneous carbon dioxide tension of 7.6 mm Hg was detectable during sedation with midazolam/alfentanil and of 2.3 mm Hg without sedation. Conclusion: Combined POX/PcCO2 monitoring at the ear lobe is a novel approach to improve patient safety during sedation and may be helpful in preventing an unintentional slide into a state of deep sedation with impairment of ventilation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.