Many recent bioanalytical systems based on immunologic and hybridization reactions in a mono- or bidimensional microarray format require technology that can produce arrays of spots containing biospecific molecules. Some microarray deposition instruments are commercially available, and other devices have been described in recent papers. We describe a system obtained by adapting a commercial ink-jet printer and used to produce mono- and bidimensional arrays of spots containing horseradish peroxidase on cellulose paper. In a few minutes, it was possible to obtain bidimensional arrays containing several thousands of spots with a diameter as low as 0.2 mm, with each of which requiring only a few nanoliters of the enzyme deposition solution. The quantity of enzyme in each spot was evaluated with a chemiluminescent reaction and a charge-coupled device-based, low-light imaging luminograph. The chemiluminescence measurements revealed that the reproducibility of the enzyme deposition was satisfactory for analytical purposes, with the variation coefficients being lower than 10% in almost all cases.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.