An x-ray dose-monitoring system using a small digital computer is described. Initially, and for every 6 months afterward, the system is calibrated using an exposure meter. For each exposure, the computer receives values of x-ray technique and beam geometry from the x-ray generator through a specially designed electronic interface. Then, by means of calibration data, entrance exposure, area exposure product, and integral dose are obtained and printed for each patient examined. The overall accuracy of the system is better than +/- 20%. Operation is semiautomatic, requiring minimum operator intervention. Over 2,000 patients have been monitored with the device. Because the system is computer-based, it offers the opportunity for statistical analysis of the data base created, as the results for each patient are stored on computer disk.
Active surveillance might serve as an alternative treatment option for older patients with PTMC. Since the median follow-up periods are currently not long enough, it seems difficult to draw definite conclusions of this procedure right now.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.