Dramatic increases in both magnetic resonance imaging (MRI) usage and cardiac device-based therapy have resulted in an estimated 50-75% probability of a patient being indicated for an MRI over the lifetime of their device. Some recent studies have demonstrated "safe procedures" and "no adverse events" in the limited populations, clinical situations, and specific devices and lead orientations tested. While these investigations are useful to help ascertain the hazards for patients with cardiac devices, they do not demonstrate clear freedom from risk. All components of active implantable systems must be engineered during the design stage to provide safety in current and evolving MR environments. Device manufacturers need to secure regulatory approval to confirm their products' safety under multiple clinical and technical variables.
A safe temperature for induced lead heating is 43 °C for 30 minutes. MRI-related RF heating above 43 °C or longer than 30 minutes may be associated with increased risk of clinically evident thermal damage to neural structures immediately surrounding implanted leads. The establishment of a thermal dose limit is a first step toward making specific neurostimulation systems conditionally safe during MRI procedures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.