The purpose of this paper is to describe our experience with the AAPM Medical Physics Practice Guideline 1.a: “CT Protocol Management and Review Practice Guideline”. Specifically, we will share how our institution's quality management system addresses the suggestions within the AAPM practice report. We feel this paper is needed as it was beyond the scope of the AAPM practice guideline to provide specific details on fulfilling individual guidelines. Our hope is that other institutions will be able to emulate some of our practices and that this article would encourage other types of centers (e.g., community hospitals) to share their methodology for approaching CT protocol optimization and quality control. Our institution had a functioning CT protocol optimization process, albeit informal, since we began using CT. Recently, we made our protocol development and validation process compliant with a number of the ISO 9001:2008 clauses and this required us to formalize the roles of the members of our CT protocol optimization team. We rely heavily on PACS‐based IT solutions for acquiring radiologist feedback on the performance of our CT protocols and the performance of our CT scanners in terms of dose (scanner output) and the function of the automatic tube current modulation. Specific details on our quality management system covering both quality control and ongoing optimization have been provided. The roles of each CT protocol team member have been defined, and the critical role that IT solutions provides for the management of files and the monitoring of CT protocols has been reviewed. In addition, the invaluable role management provides by being a champion for the project has been explained; lack of a project champion will mitigate the efforts of a CT protocol optimization team. Meeting the guidelines set forth in the AAPM practice guideline was not inherently difficult, but did, in our case, require the cooperation of radiologists, technologists, physicists, IT, administrative staff, and hospital management. Some of the IT solutions presented in this paper are novel and currently unique to our institution.PACS number: 87.57.Q
This article explains a method for creating CT protocols for a wide range of patient body sizes and clinical indications, using detailed tube current information from a small set of commonly used protocols. Analytical expressions were created relating CT technical acquisition parameters which can be used to create new CT protocols on a given scanner or customize protocols from one scanner to another. Plots of mA as a function of patient size for specific anatomical regions were generated and used to identify the tube output needs for patients as a function of size for a single master protocol. Tube output data were obtained from the DICOM header of clinical images from our PACS and patient size was measured from CT localizer radiographs under IRB approval. This master protocol was then used to create 11 additional master protocols. The 12 master protocols were further combined to create 39 single and multiphase clinical protocols. Radiologist acceptance rate of exams scanned using the clinical protocols was monitored for 12,857 patients to analyze the effectiveness of the presented protocol management methods using a two‐tailed Fisher's exact test. A single routine adult abdominal protocol was used as the master protocol to create 11 additional master abdominal protocols of varying dose and beam energy. Situations in which the maximum tube current would have been exceeded are presented, and the trade‐offs between increasing the effective tube output via 1) decreasing pitch, 2) increasing the scan time, or 3) increasing the kV are discussed. Out of 12 master protocols customized across three different scanners, only one had a statistically significant acceptance rate that differed from the scanner it was customized from. The difference, however, was only 1% and was judged to be negligible. All other master protocols differed in acceptance rate insignificantly between scanners. The methodology described in this paper allows a small set of master protocols to be adapted among different clinical indications on a single scanner and among different CT scanners.PACS number: 87.57.Q
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.