Retention behavior of U(VI) and Th(IV) as their 2,6-pyridine dicarboxylic acid (PDCA) complexes on reversed phase and ion exchange (cation, anion and mixed ion exchange) columns was studied and based on the results, a simple ion chromatography method for the determination of trace level thorium in uranium oxide using 0.075 mM 2,6-pyridine dicarboxylic acid (PDCA) and 1 M KNO 3 in 1.2 M HNO 3 as eluent (flow rate 1 mL/min) was proposed. The advantage of the developed method is that the separation of uranium matrix is not required prior to the ion chromatographic determination of trace Th. Separation was carried out on a mixed ion exchange stationary phase and a 10 −4 M arsenazo (III) solution was used as post column reagent for detecting the separated metal ions. The separation of Th from uranium using PDCA in the present investigation is attributed through cation exchange mechanism. A calibration plot was constructed by following the standard addition method over the concentration range of 0.25 to 10 ppm of Th in the presence of uranium matrix, which resulted in a linear regression coefficient of 0.9978. The precision of the method was better than 5% and the LOD for Th was found to be 0.1 ppm (S/N = 3). The method has been validated by comparing the results with the results obtained from ICP-MS analysis where the Th is separated from the uranium matrix. The proposed method is simple, rapid, accurate and cost effective compared to techniques like ICP-MS or ICP-AES and is suitable for the routine kind of analysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.