After preparing an all-solid-state copper ion selective electrode (Cu 2+ -ISE) based on PEDOT/PSS conversion layer via electrodeposition, this paper characterizes the morphology and cyclic voltammetry performance of the PEDOT/PSS film, optimizes the deposition time of the PEDOT/PSS film, and tests the reproducibility, selectivity and acid-base resistance of the all-solid-state copper ion selective electrode. The experimental results show that the prepared all-solid-state copper ion selective electrode has a linear response in the range of 1×10 -6 mol•L -1 ~1×10 -2 mol•L -1 , with a response time less than 30 s, and the response slope of 27.8 mV/ decade. The all-solid-state copper ion selective electrode mentioned above can meet the detection standard of national II surface water.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.