Patients with less severe obstructive sleep apnoea (OSA) are usually prescribed oral appliances and/or smaller optimal nasal continuous positive airway pressure (PnCPAP) in nCPAP therapy. We hypothesised that OSA patients with greater PnCPAP would not respond favourably to oral appliances.Oral appliances were inserted in nCPAP users after washing-out the nCPAP effect. Follow-up polysomnography was undertaken with the adjusted oral appliance in place. The predictability of PnCPAP was evaluated with receiver-operating characteristic (ROC) curves.The median baseline apnoea/hypopnoea index (AHI) was reduced with the oral appliance from 36 to 12 events?h -1 in 35 patients. When responders were defined as patients showing a follow-up AHI of ,5 events?h -1 with .50% reduction in baseline AHI, the area under the ROC curve for PnCPAP was 0.76. The best cut-off value of PnCPAP turned out to be 10.5 cmH 2 O with a high negative predictive value (0.93) and a low negative likelihood ratio (0.18). OSA patients with a PnCPAP of .10.5 cmH 2 O are unlikely to respond to oral appliance therapy. This prediction is clinically helpful to both OSA patients and medical personnel in discussing oral appliances as a temporary substitute and/or alternative for nCPAP.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.