This study presents the use of molecularly imprinted polymer (MIP) as packing material for microextraction by packed syringe (MEPS) to achieve higher extraction selectivity. Pentycaine was used as template for MIP. Development and validation of the determination of lidocaine, ropivacaine, mepivacaine and bupivacaine in human plasma and urine samples utilizing MIP-MEPS and liquid chromatography-tandem mass spectrometry (LC-MS/MS) were carried out. The MEPS MIP-cartridge could be used for 100 extractions before it was discarded. The extraction recovery ranged from 60 to 80%. The correlation coefficients values were >0.999 for all assays using lidocaine, ropivacaine, mepivacaine and bupivacaine in the calibration range 5-2000 nmol/L. The accuracy of the studied compounds, given as a percentage variation from the nominal concentration values, ranged from -4.9 to 8.4% using plasma and urine samples. The between-batch precision, given as the relative standard deviation, at three different concentrations (quality control samples) was ranged from -4.7 to 14.0% and from 1.8 to 12.7% in plasma and urine, respectively. The lower limit of quantification and limit of detection of the studied substances were 5.0 and 1.0 nm, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.