We consider the eight-Plate to be an ideal tool for treatment of angular deformities in growing children. It allows for precise insertion and is reliable. It is also less likely to extrude like the Blount staple.
Background Patients with congenital limb shortening can present with joint instability, soft tissue contractures, and significant leg length discrepancy. Classically, lengthening is done with external fixation, which can result in scarring, pin site infection, loss of motion, and pain. We therefore developed an alternative to this approach, a new, controllable, internal lengthening device for patients with congenital limb shortening.
Level IV-therapeutic.This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal. http://creativecommons.org/licenses/by-nc-nd/4.0/.
Introduction: Femoral/tibial lengthening with a telescopic, magnetically-powered, intramedullary nail is an alternative to lengthening with external fixation.
Methods: Pre-clinical testing was conducted of the PRECICE in a human cadaver. A retrospective review of the first 30 consecutive patients who underwent unilateral lengthening was also conducted. Nail accuracy was obtained by comparing the amount of nail distraction to the final bone length achieved at the end of the distraction process. Relative standard deviation of accuracy was used to calculate nail precision.
Results: Devices performed successfully in a human cadaver. Thirty consecutive patients (10 females, 20 males; mean age, 23 years) with limb length discrepancy (LLD) were followed an average of 19 months (range, 12–24 months). Etiology included congenital shortening (14), posttraumatic deformities (7), Ollier disease (3), osteosarcoma resection (1), prior clubfoot (2), hip dysplasia (1), post-septic growth arrest of knee (1), and LLD after hip arthroplasty (1). Twenty-four femoral and eight tibial nails were implanted. Mean preoperative lengthening goal was 4.4 cm (range, 2–6.5 cm); mean postoperative length achieved was 4.3 cm (range, 1.5–6.5 cm). Average consolidation index was 36.4 days/cm (range, 12.8–113 days/cm). Mean nail accuracy was 97.3% with a precision of 92.4%. Average preoperative and 12-month postoperative Enneking scores were 21.5 and 25.3 (p < 0.001), respectively. The preoperative and 12-month postoperative SF-12 physical and mental component scores were not statistically different. Nine complications (nine limb segments) resolved: two partial femoral unions, two suspected deep vein thrombosis (DVT), one delayed tibial union, one fibular nonunion, one peroneal nerve irritation, one knee joint subluxation, and one confirmed DVT. Twenty-nine (91%) of 32 limb segments achieved successful bone healing without revision surgery.
Discussion: Limb lengthening with PRECICE is reliable, but larger trials with longer follow-up will reveal limitations. Implantable nails prevent problems associated with external fixation, such as muscle tethering and pin-site infections.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.