Percutaneous cardiopulmonary support (PCPS) has come to be applied for cardiopulmonary resuscitation and in the management of severe respiratory failure as well as severe heart failure. We investigated cerebral tissue oxygen saturation during PCPS in a canine model of respiratory failure using near-infrared spectroscopy. Animals were mechanically ventilated with 10% oxygen to make a respiratory failure model. Perfusion with PCPS was performed via the left femoral artery and switched to that via the right axillary artery. Cerebral tissue oxygen saturation was 54.2 +/- 3.4% during PCPS via the femoral artery and was 82.3 +/- 4.6% during PCPS via the axillary artery (p = 0.001). Hepatic tissue oxygen saturation was not significantly different. LV dP/dt max increased significantly after switching to the axillary blood supply (p = 0.001). Conventional PCPS may not have the capability of supporting cerebral circulation under severe respiratory failure without organic heart disease.
A computer simulation was carried out to investigate the influence of nonpulsatile left ventricular assistance on hemodynamics. A simulation circuit was constructed to represent the circulatory system. A source of current was added to denote the nonpulsatile blood pump. The left and right ventricles were replaced by variable compliances. Left heart failure was simulated by decreasing the amount of compliance change of the left ventricle. We introduced a pulsatility indicator (PI) to clarify the pulsatility characteristics in the hemodynamics; this PI was defined as the ratio of the pulse pressure (PP) to the mean aortic pressure (AoP). When nonpulsatile bypass flow increased, the mean AoP, tension time index (TTI), and diastolic pressure time index (DPTI) increased, and cardiac output, PP, and PI decreased. When assisted flow increased with the constant total flow rate, the mean AoP and DPTI changed little; the PP, TTI, and PI decreased, and the endocardial viability rate increased. The PI would be helpful in evaluating the effect of pulsatility.
In order to develop a diaphragm-type ventricular assist device (VAD), we studied the flow field change following structural modifications. We devised a center flow-type pump by putting a small projection on the center of the housing and/or diaphragm to provide a center in the flow field, and examined the following four types of VADs: N type without a projection, D type with a projection on the diaphragm, H type with a projection on the housing, and DH type with projections on both the diaphragm and housing. Computational fluid dynamics (CFD) was used for flow simulation. Particle image velocimetry (PIV) was also used to verify the reliability of the CFD method and to determine how the flow field changes in the presence of a projection. The results of the PIV and CFD analyses were comparable. The placement of a projection on the housing was most effective in rectifying the flow field.
Working toward a completely implantable total artificial heart, we have designed an eccentric roller type total artificial heart. The actuator of this artificial heart is a drum type eccentric roller that squeezes the blood chambers. The blood chambers are made of silicone rubber and are torus in shape. The shape of the artificial heart is an almost circular cylinder, and its length and diameter are 10 cm and 8 cm, respectively. The 2 main characteristics of this artificial heart are that it discharges blood in a pulsatile mode and that it requires no reversing of the motor. Because we have not completed the artificial heart yet, we have tested the eccentric roller mechanism on the prototype with an overflow type mock circulation with a 100 mm Hg afterload. The prototype worked at the roller speeds of 50, 100, and 150 rpm with flow rates of 1.7, 3.7, and 5.4 L/min, respectively. Next the prototype was connected to a Donovan type mock circulatory system and worked at roller speeds of 88-214 rpm with flow rates of 3.0-8.4 L/min against mean afterloads of 82-120 mm Hg.
The influence of the Wankel type semipulsatile left ventricular assistance on hemodynamics was investigated with a computer simulation and an animal experiment. A simulation circuit was constructed to express the circulatory system. A current source was added to create a semipulsatile blood pump. The left and right ventricles were replaced by variable compliances. Left heart failure was simulated by decreasing the amount of compliance change of the left ventricle. Under the condition of heart failure when semipulsatile assist flow increased, the mean aortic pressure (AoP), tension time index (TTI), and diastolic pressure time index (DPTI) increased, and the cardiac output, pulse pressure (PP), and pulsatility indicator (PI) decreased. In an animal experiment, a Wankel type blood pump was used in a calf. With the increase of the assist flow, AoP curves became less pulsatile, and PP and PI decreased in accordance, which was predicted by the numerical simulation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.