Axillary vein access using fluoroscopic landmarks, especially the body surface of the second rib, is an effective approach for ICD implantation and offers the potential to avoid complications usually observed with traditional subclavian vein approach.
RBBB is the most common IVCD. Its presence can be explained by a susceptibility of the right branch to heart positioning, pressure overload, and acute rejection. IVCDs do not affect prognosis.
Aims
Pocket hematoma is a common complication of cardiac implantable electronic device (CIED) procedures. the aim of the study was to research the clinical factors associated with pocket hematoma formation after CIED implantation or replacement and to identify the best perioperative antithrombotic management.
Methods
We retrospectively analyzed 500 consecutive patients who underwent to CIED implantation or replacement at our center from November 2014.
Results
Among our population, 206 patients (41.2%) were on anticoagulant therapy at the time of the intervention: 68 (13.6%) on ongoing Warfarin; 111 (22.2%) on low-molecular-weight heparin (LMWH); and 27 (5.4%) on ongoing direct oral anticoagulants. Antiplatelet therapy was present in 262 (52.4%) patients: in particular, 50 (10%) were on dual antiplatelet therapy, 64 (12.8%) were on single antiplatelet therapy and anticoagulant therapy, whereas 12 (2.4%) were on anticoagulant with dual antiplatelet therapy.
Incidence of pocket hematoma after CIEDs implantation was of 4.6%. Considering the different perioperative anticoagulant strategies, patients on LMWH presented the higher hematoma rate [11/100 patients (11.0%), P < 0.001]. At the multivariate analysis, anticoagulant with dual antiplatelet therapy (P = 0.021, OR 6.3, IC 1.3–30.8), left ventricular ejection fraction (LVEF) less than 30% (P < 0.001, OR 7.4, IC 2.7–20.4), and use of LMWH (P = 0.008, OR 3.8, IC 1.4–10.6) resulted the strongest predictors of pocket hematoma (Hosmer test = 0.899).
Considering replacement procedures, incidence of pocket hematoma was of 4.4%. The incidence was higher after ICD/CRT-D replacement. The majority of pocket hematoma occurred in patients with mechanical valve prosthesis (3/4 cases, 75%, P < 0.001).
Conclusion
The use of LMWH and a low LVEF expose patients to a higher risk of pocket hematoma after CIED procedures. Anticoagulant with dual antiplatelet therapy and LMWH should be avoided.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.