In this study, an innovative fast X-ray photon-counting pixel for high X-ray flux applications is proposed. A computed tomography system typically uses X-ray fluxes up to 108 photons/mm2/sec at the detector and thus a fast read-out is required in order to process individual X-ray photons. Otherwise, pulse pile-up can occur at the output of the signal processing unit. These superimposed signals can distort the number of incident X-ray photons leading to count loss. To minimize such losses, a cross detection method was implemented in the photon-counting pixel. A maximum count rate under X-ray tube voltage of 90 kV was acquired which reflect electrical test results of the proposed photon counting pixel. A maximum count of 780 kcps was achieved with a conventional photon-counting pixel at the pulse processing time of 500 ns, which is the time for a pulse to return to the baseline from the initial rise. In contrast, the maximum count of about 8.1 Mcps was achieved with the proposed photon-counting pixel. From these results, it was clear that the maximum count rate was increased by approximately a factor 10 times by adopting the cross detection method. Therefore, it is an innovative method to reduce count loss from pulse pile-up in a photon-counting pixel while maintaining the pulse processing time.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.