The Iseult/INUMAC 11.7 T MRI magnet uses NbTi conductors larger than those typically used in conventional MRI magnets. The principal conductors are the main coil one, producing the magnetic field of 11.7 T and the conductor of the shielding coil that limits the fringe field. Both conductors are being produced by Luvata Waterbury Inc. The main coil conductor is a Rutherford cable in a copper channel (160 km) and carries 1500 A at 12 T and 2.8 K (2727 A at 9.5 T and 4.2 K). The shielding coil conductor is a wire in channel (60 km) and carries 2100 A at 5 T and 4.2 K. The conductor shape must be very precise with reproducible dimensions (tolerances below 15 microns) to enable good magnetic field homogeneity. They must also simultaneously exhibit a low electrical resistance (20 at 10 K and 12 T) and a high mechanical strength . Conductors of this industrial production are characterized by mechanical, electrical and geometrical measurements.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.