Zeptomole detector: A highly sensitive giant-magnetoresistive chip and FeCo nanoparticles can be used to linearly detect 600-4500 copies of streptavidin. Under unoptimized conditions, this system also detects human IL-6 with a sensitivity 13-times higher than that of standard ELISA techniques.
A novel giant magnetoresistive sensor and uniform high-magnetic-moment FeCo nanoparticles (12.8 nm)-based detecting platform with minimized detecting distance was developed for rapid biomolecule quantification from body fluids. Such a system demonstrates specific, accurate, and quick detection and quantification of interleukin-6, a low-abundance protein and a potential cancer biomarker, directly in 4 muL of unprocessed human sera. This platform is expected to facilitate the identification and validation of disease biomarkers. It may eventually lead to a low-cost personal medical device for chronic disease early detection, diagnosis, and prognosis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.