The transient hot-wire (THW) technique is an absolute technique for measuring the thermal conductivity of fluids. In this work, a THW apparatus was developed for measurements over the temperature range from (193 to 424) K at pressures to 34 MPa. The apparatus was commissioned with the measurements of pure argon, methane and propane along several isotherms at pressures to 33 MPa. The measured values agreed with those calculated with the reference equations within the uncertainty of the equations. Thermal conductivity measurements of a binary mixture (0.9514 methane + 0.0486 propane) were then carried out in the temperature range from (195 to 424) K at pressures to 31 MPa in the single phase and near the bubble point curve. The relative combined expanded uncertainty (k = 2) for these single-phase and bubble-point mixture measurements was estimated as 2.0 % and 5.0 %, respectively. The relative deviations of these measured values from those calculated with the extended corresponding states model implemented in the software package REFPROP 10.0 were within 4.4 % and 8.0 %, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.