Calibration of the Wireless Unit of Micromechanical Accelerometers

Abstract: The report describes an original method of calibrating an autonomous unit of wireless accelerometers manufactured by Microdevices Laboratory, Ltd [1]. Without the use of precision rotary tables, centrifuges, and an independent basement (against vibrations), it is possible to reduce the error caused by the mismatch (not orthogonal)of sensor’s axes. Zero and scale factors of the sensor unit are calibrated with great precision. The calibration of the sensor unit with a measuring range of 2 g, produces the followi… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2013
2013
2021
2021

Publication Types

Select...
1
1

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(1 citation statement)
references
References 0 publications
0
1
0
Order By: Relevance
“…The angle between sensor A and B is 90 degree. Turntable is rotated one circle clockwise and counterclockwise, respectively [3][4] . Then, the accelerometers output values are written down.…”
Section: Framework For the Testing Systemmentioning
confidence: 99%
“…The angle between sensor A and B is 90 degree. Turntable is rotated one circle clockwise and counterclockwise, respectively [3][4] . Then, the accelerometers output values are written down.…”
Section: Framework For the Testing Systemmentioning