As a widely clinical detection method, ultrasonography (US) has been applied to the diagnosis of breast cancer. In this paper, the multi-element synthetic aperture focusing (M-SAF) is applied to the ring array of breast ultrasonography (US) imaging, which addresses the problem of low imaging quality due to the single active element for each emission and the reception in the synthetic aperture focusing. In order to determine the optimal sub-aperture size, the formula is derived for calculating the internal sound pressure of the ring array with a 200 mm diameter, and the sound pressure distribution is analyzed. The ring array with 1024 elements (1024 ring array) is established in COMSOL Multiphysics 5.6, and the optimal sub-aperture size is 16 elements, according to the sound field beam simulation and the directivity research. Based on the existing experimental conditions, the ring array with 256 elements (256 ring array) is simulated and verified by experiments. The simulation has a spatial resolution evaluation in the k-Wave toolbox, and the experiment uses nylon rope and breast model imaging. The results show that if the sub-aperture size has four elements, the imaging quality is the highest. Specifically, the spatial resolution is the best, and the sound pressure amplitude and signal-to-noise ratio (SNR) are maintained at a high level in the reconstructed image. The optimal sub-aperture theory is verified by the two kinds of ring arrays, which also provide a theoretical basis for the application of the multi-element synthetic aperture focusing technology (M-SAF) in ring arrays.