Dual-Energy Low-keV or Single-Energy Low-kV CT for Endoleak Detection?

Abstract: The aim of this study was to compare image quality, conspicuity, and endoleak detection between single-energy low-kV images (SEIs) and dual-energy low-keV virtual monoenergetic images (VMIs+) in computed tomography angiography of the aorta after endovascular repair. MATERIALS AND METHODS: An abdominal aortic aneurysm phantom simulating 36 endoleaks (2 densities; diameters: 2, 4, and 6 mm) in a medium-and large-sized patient was used. Each size was scanned using single-energy at 80 kVp (A) and 100 kVp (B), and … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

1
12
1

Year Published

2020
2020
2024
2024

Publication Types

Select...
7
1

Relationship

3
5

Authors

Journals

citations
Cited by 18 publications
(14 citation statements)
references
References 35 publications
1
12
1
Order By: Relevance
“…Our results for objective image quality are comparable to a previous phantom study comparing advanced VMIs (VMI+) of dual-energy EID-CT to single-energy low-kV imaging EID-CT on a third-generation dual-source CT system. 35 Here, CNR was significantly higher for VMI at 40 and 50 keVas compared with single-energy at 80 kV. As a difference to this study, in which subjective image quality was significantly lower for VMI, we did not find a significant difference in subjective overall image quality among 40 keV, 45 keV, and EID-CT.…”
Section: Discussioncontrasting
confidence: 85%
See 1 more Smart Citation
“…Our results for objective image quality are comparable to a previous phantom study comparing advanced VMIs (VMI+) of dual-energy EID-CT to single-energy low-kV imaging EID-CT on a third-generation dual-source CT system. 35 Here, CNR was significantly higher for VMI at 40 and 50 keVas compared with single-energy at 80 kV. As a difference to this study, in which subjective image quality was significantly lower for VMI, we did not find a significant difference in subjective overall image quality among 40 keV, 45 keV, and EID-CT.…”
Section: Discussioncontrasting
confidence: 85%
“…Our results for objective image quality are comparable to a previous phantom study comparing advanced VMIs (VMI+) of dual-energy EID-CT to single-energy low-kV imaging EID-CT on a third-generation dual-source CT system 35 . Here, CNR was significantly higher for VMI at 40 and 50 keV as compared with single-energy at 80 kV.…”
Section: Discussionsupporting
confidence: 82%
“…The CTA at low tube voltages (low-kV imaging) has been investigated as a tool to increase CT attenuation and the contrast-to-noise ratio (CNR) of the aorta with the option of radiation dose and/or CM reduction 18–30 . Following these lines, automatic tube voltage selection (ATVS) has been introduced as a means to reduce radiation dose in CT depending on the imaging task and patient habitus, while maintaining image quality in terms of the CNR 31 .…”
mentioning
confidence: 99%
“…17 The CTA at low tube voltages (low-kV imaging) has been investigated as a tool to increase CT attenuation and the contrast-to-noise ratio (CNR) of the aorta with the option of radiation dose and/or CM reduction. [18][19][20][21][22][23][24][25][26][27][28][29][30] Following these lines, automatic tube voltage selection (ATVS) has been introduced as a means to reduce radiation dose in CT depending on the imaging task and patient habitus, while maintaining image quality in terms of the CNR. 31 The ATVS system has primarily been designed for radiation dose optimization by selecting the tube voltage level that offers lowest radiation dose at maintained CNR.…”
mentioning
confidence: 99%
“…Endoleaks are associated with an increased aneurysm rupture rate and a reduced viability and patency of the endograft itself 4 . Therefore, in patients treated with EVAR, life-long follow-up imaging, usually with multiphasic computed tomography (CT) angiography, is necessary 5,6 . The typical CT protocol consists of a true noncontrast (TNC) phase, an arterial phase, and a venous phase 5 .…”
mentioning
confidence: 99%