2003
DOI: 10.1046/j.1460-9592.2003.t01-1-00183.x
|View full text |Cite
|
Sign up to set email alerts
|

Electrical Interference from an Abdominal Muscle Stimulator Unit on an Implantable Cardioverter Defibrillator:

Abstract: Two consecutive cases are presented of patients with ICDs in whom the use of commercial units for muscle stimulation for abdominal training caused interference with the device, mimicked cardiac signals, and resulted in inappropriate treatment shock delivery.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
8
0

Year Published

2006
2006
2019
2019

Publication Types

Select...
4
2

Relationship

0
6

Authors

Journals

citations
Cited by 13 publications
(8 citation statements)
references
References 4 publications
0
8
0
Order By: Relevance
“…reported on two male patients with pectoral implanted ICDs (Ventak Mini III® and Ventak AV III DR®) and neuromuscular stimulation of abdominal muscles [33]. In both patients, NMES using commercial units (5.0–11.0 V, 7.3–10 mA, 55–75 Hz/biphasic) provoked EMI, resulting in ICD discharge due to misinterpreting electrical signals as cardiac signals in the ventricular fibrillation (VF) zone (Table 1; [33]).…”
Section: Resultsmentioning
confidence: 99%
See 3 more Smart Citations
“…reported on two male patients with pectoral implanted ICDs (Ventak Mini III® and Ventak AV III DR®) and neuromuscular stimulation of abdominal muscles [33]. In both patients, NMES using commercial units (5.0–11.0 V, 7.3–10 mA, 55–75 Hz/biphasic) provoked EMI, resulting in ICD discharge due to misinterpreting electrical signals as cardiac signals in the ventricular fibrillation (VF) zone (Table 1; [33]).…”
Section: Resultsmentioning
confidence: 99%
“… Home therapy (NMES) of thighs : no EMIWayar et al. 2003 [33]Case report ( n  = 2)1st patient: pectoral ICD (Ventak®  Mini III [Guidant Inc., St. Paul, Minnesota, USA])NMES of abdominal muscles55–75 Hz, biphasic, 7.3–10 mA, 5–11 VEMI in both patients2nd patient: pectoral ICD (Ventak® AV [Guidant Inc., St. Paul, Minnesota, USA]) ICD  implantable cardioverter defibrillator, NMES  neuromuscular electrical stimulation, stim  stimulation, EMI  electromagnetic interference, IG50  Impulse galvanization, dir. cur.…”
Section: Resultsmentioning
confidence: 99%
See 2 more Smart Citations
“…[1][2][3] The majority of these shocks (>80%) stem from nonsustained VTs or supraventricular tachycardias. Electromagnetic interference accounts for a minority of inappropriate shocks (<5%) and has been shown to arise from normally [4][5][6][7][8] and abnormally 9-12 functioning electrical appliances.…”
Section: Discussionmentioning