Abstract:This study aimed to verify the accuracy of half‐value layer (HVL) measured using the new copper pipe method with the CT ionization chamber while the X‐ray tube is rotating and to compare it with the conventional nonrotating method and Monte Carlo simulation method based on the actual measurement and geometry of the new copper pipe method. HVL was measured while the X‐ray tube was rotating using a CT ionization chamber surrounded by copper pipe absorbers and located at the isocenter of the CT gantry. The exposu… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.