Order By: Relevance
“…260-261 of Ref. 40) and found an effective gain = 5.13 × 10 −9 volts per optical secondary. To model the quantum noise in a pixel, we invoked the noise-free gain assumption so that the quantum noise variance of g m may be written as the mean number of absorbed x rays times the total gain squared (pp.…”
Section: Iic Detector Modelmentioning
confidence: 99%
See 1 more Smart Citation
“…260-261 of Ref. 40) and found an effective gain = 5.13 × 10 −9 volts per optical secondary. To model the quantum noise in a pixel, we invoked the noise-free gain assumption so that the quantum noise variance of g m may be written as the mean number of absorbed x rays times the total gain squared (pp.…”
Section: Iic Detector Modelmentioning
confidence: 99%
“…CsI:Tl has a gain of 55 optical photons per keV of incident x-ray energy. 40 For 20 keV photons, we assumed η(20 keV) ≈ 70% so ηk ≈ 800 optical secondaries per x-ray primary. The electronic gain parameter was calculated from Eq.…”
Section: Iic Detector Modelmentioning
confidence: 99%
“…For the last two decades amorphous Silicon and amorphous Selenium based flat panel imagers (FPI) have represented the detector of choice in digital medical imaging, mainly because they can be easily fabricated on a large area with a relatively low cost technology, derived from the consumer based flat panel display technology [1].…”
Section: Introductionmentioning
confidence: 99%
“…amorphous Silicon (a-Si) or amorphous Selenium (a-Se) thin film transistors (TFT)) (Rowlands and Yorkston 2000). This is especially common in medical imaging applications where a large imaging area is required (≥ 43 cm × 43 cm for chest radiography, 25 cm × 30 cm for mammography, 15 − 30 cm radius for fluoroscopy, ≥ 41 cm × 41 cm for radiotherapy portal imaging), since they rely on a 1:1 aspect ratio covering the entire imaging area.…”
Section: Introductionmentioning