Background: Pulsatile perfusion during extracorporeal circulation is a promising concept to improve perfusion of critical organs. Clinical benefits are limited by the amount of pulsatile energy provided by standard pumps. The present study investigated the properties of a novel positive displacement blood pump in a mock circulation.
Methods:The pump was attached to an aortic model with a human-like geometry and compliance as a pseudo patient. Hemodynamic data were recorded while the pump settings were adjusted systematically.Results: Using a regular oxygenator, maximum flow was 2.6 L/min at a pressure of 27 mm Hg and a frequency (F) of 90 bpm. Pulse pressure (PP; 28.9 mm Hg) and surplus hemodynamic energy (SHE; 26.1% of mean arterial pressure) were highest at F = 40 bpm. Flow and pressure profiles appeared sinusoid. Using a lowresistance membrane ventilator to assess the impact of back pressure, maximum flow was 4.0 L/min at a pressure of 58.6 mm Hg and F = 40 bpm. At F = 40 bpm, PP was 58.7 mm Hg with an SHE of 33.4%. SHE decreased with increasing flow, heart rate, and systolic percentage but surpassed 10% with reasonable settings.