Infusion technology: a cause for alarm

Abstract: Occlusion pressure limit alarm cannot reliably detect extravasation especially at sites with high compliance, low flow rates, even at low occlusion limit. Line pressure depends on interplay of site (compliance) and flow rate and is independent of volume extravasated. Users must be aware of the set occlusion pressure limit. Repeated clinical assessment remains vital.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4

Citation Types

0
15
0
1

Year Published

2005
2005
2022
2022

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 14 publications
(16 citation statements)
references
References 16 publications
0
15
0
1
Order By: Relevance
“…Third, the use of a pressurized infusion pump contributes to the development of compartment syndrome when extravasation occurs 3) . The pressurized infusion pump is a medical device that applies pressure at a consistent speed to inject fluid, and measures the fluid collected from the chamber of the intravenous infusion set 9) . The pressurized infusion pump is currently used extensively for pediatric patients whose volume of distribution is small and who require a more precise adjustment of the rate of fluid injection 9) .…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Third, the use of a pressurized infusion pump contributes to the development of compartment syndrome when extravasation occurs 3) . The pressurized infusion pump is a medical device that applies pressure at a consistent speed to inject fluid, and measures the fluid collected from the chamber of the intravenous infusion set 9) . The pressurized infusion pump is currently used extensively for pediatric patients whose volume of distribution is small and who require a more precise adjustment of the rate of fluid injection 9) .…”
Section: Discussionmentioning
confidence: 99%
“…The pressurized infusion pump is a medical device that applies pressure at a consistent speed to inject fluid, and measures the fluid collected from the chamber of the intravenous infusion set 9) . The pressurized infusion pump is currently used extensively for pediatric patients whose volume of distribution is small and who require a more precise adjustment of the rate of fluid injection 9) . Despite this advantage, it has a risk of infusing fluid continuously even when the internal pressure increases because of extravasation in an enclosed space 3) .…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…The linear peristaltic pump has a row of horizontally placed fingers that sequentially compress the intravenous tubing in a wavelike fashion [11,12]. The accuracy of those devices is usually within 5-10% of the selected flow rate [13,14]. The Terufusion TE-112 and Volumed μVP7000 used in the present study are microprocessor-controlled volumetric pumps that use a peristaltic system.…”
Section: Discussionmentioning
confidence: 99%
“…This provides a clinical warning of flow faults caused by excessive backpressure downstream in the line. 2 Often, however, fluid leakage fails to yield pressures that would trigger an alarmed state even with low occlusion limit settings, in part, due to low flow rates and skin compliance. 2Y4 It has also been clearly documented that occlusion alarms are not intended to detect catheter infiltration or fluid extravasation, 2,3 although it has been suggested that monitoring flow resistance (rather than pressure) may be a more appropriate method for early detection of these issues.…”
mentioning