“…Limits of detection and quantitation. The limit of detection (LOD) and limit of quantitation (LOQ) were estimated statistically from a diluted calibration curve (0.01-1.60 mg/mL of AA), as 3*SE/S and 10*SE/S respectively (Ermer et al, 2005;Hibbert, 2007), where SE is the standard error of the regression and S the slope of the curve. In addition, the LOD and LOQ were estimated empirically, by progressively measuring AA in serially diluted AA standard solutions (0.01-0.2 mg/mL).…”