Quantifying Physical Thrombus Characteristics on Cardiovascular Biomaterials Using MicroCT

Abstract: Hemocompatibility is a critical consideration when designing cardiovascular devices. Methods of assessing hemocompatibility range from in vitro protein adsorption and static platelet attachment to in vivo implantation. A standard preclinical assessment of biomaterial hemocompatibility is ex vivo quantification of thrombosis in a chronic arteriovenous shunt. This technique utilizes flowing blood and quantifies platelet accumulation and fibrin deposition. However, the physical parameters of the thrombus have rem… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
7
0

Year Published

2020
2020
2023
2023

Publication Types

Select...
4

Relationship

3
1

Authors

Journals

citations
Cited by 4 publications
(7 citation statements)
references
References 30 publications
0
7
0
Order By: Relevance
“…Additional controls were unpatterned, planar PVA tubes (4 mm ID, n = 2), clinical ePTFE grafts (4 mm ID, n = 2), and collagen-coated ePTFE grafts (4 mm ID, n = 2), which served as negative, clinical, and positive controls, respectively. These samples were not compared to the micropatterned grafts statistically, but results were comparable with prior historical controls ( Cutiongco et al, 2015 ; Anderson et al, 2019 ; Bates et al, 2020 ; Gupta et al, 2020 ; Yao et al, 2020 ).…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…Additional controls were unpatterned, planar PVA tubes (4 mm ID, n = 2), clinical ePTFE grafts (4 mm ID, n = 2), and collagen-coated ePTFE grafts (4 mm ID, n = 2), which served as negative, clinical, and positive controls, respectively. These samples were not compared to the micropatterned grafts statistically, but results were comparable with prior historical controls ( Cutiongco et al, 2015 ; Anderson et al, 2019 ; Bates et al, 2020 ; Gupta et al, 2020 ; Yao et al, 2020 ).…”
Section: Methodsmentioning
confidence: 99%
“…After 24–72 h of fixation, samples were rinsed and processed for micro-computed tomography (microCT) imaging. Tube lumens were filled with Microfil ® (Flow Tech, Inc.) and cured for 2 h at room temperature to render the tube lumens radiopaque, as described previously ( Gupta et al, 2020 ). Positive and clinical control ePTFE grafts were soaked in Lugol’s solution (Sigma) for 16 h to render the thrombus radiopaque.…”
Section: Methodsmentioning
confidence: 99%
“…Some of the samples used in the shunt studies were examined with micro-computed tomography (MicroCT). The samples were prepared as described previously (Gupta et al, 2020). Briefly, PVA samples were filled with Microfil R polymer to render the lumens radiopaque.…”
Section: Whole Blood Platelet and Fibrin Adhesion Under Flow Quantificationmentioning
confidence: 99%
“…Micro-computed tomography (Inveon, Siemens) imaging was used to analyze thrombus area over the length of the samples (volume) in the small-diameter vascular grafts formed under flow by whole blood in ex vivo shunts, as described previously (Gupta et al, 2020). Imaging was performed with 2 × 2 binning, 220 projections, at an exposure of 660 ms/projection.…”