Background-18 F-Fluorodeoxyglucose (FDG) positron emission tomography/computed tomography (PET/CT) has been recently acknowledged as a diagnostic tool for prosthetic valve endocarditis, but its specificity is limited by uptake on noninfected valves. The objective of this study was to outline the main features of FDG uptake on PET/CT in patients with noninfected prosthetic heart valve (PHV). Methods and Results-Our institution's PET/CT database was reviewed to identify patients with PHV, excluding those suspected of infection or who had received antibiotic treatment. PET indication, valve location, and type (biological/ mechanical) and time from implantation were collected for each patient. Images with and without attenuation correction were considered for interpretation. The pattern of FDG uptake (absent, homogeneous, or heterogeneous) was recorded. Fifty-four PHVs (51 patients) were identified, including 32 biological valves. Indications for PET were oncology (n=26), suspicion of prosthetic valve endocarditis subsequently excluded (n=17), and history of vasculitis (n=11). A periprosthetic FDG uptake was present in 47 (87%) and 30 (56%) PHVs with and without attenuation correction, respectively, and the pattern was homogeneous in all but 4 (7%) and 3 (6%) PHVs, respectively. On quantitative analysis, maximum standardized uptake values was greater in mechanical than in biological valves