“…Compared with a traditional glass tube RE, the nanochannel array of the novel RE can signicantly decrease the loss of chloride ions in the KCl solution and effectively increase the service life of the RE. This shows that the dri rate of a RE reference potential is proportional to the diameter of the ion exchange channel, as obtained by Zevenbergen et al 23 The Ag/AgCl electrode with a thickness of about 0.5 mm was fabricated on a Pyrex 7740 glass using a silver sputtering li-off process and electrochemical treatment using a HCl solution. A hole with a 1.4 mm diameter was prepared using a laser for injecting a saturated KCl solution.…”
Section: Miniaturized Ag/agcl Re Designsupporting
confidence: 68%
“…A typical glass tube RE uses a semipermeable membrane to exchange the internal solution with the external water environment. Previous studies have used high-density microporous polyethylene 22 with a single micron hole 23 instead of the semipermeable membrane of the traditional glass tube Ag/AgCl electrode. This study gradually improved the performance of the electrode, especially the life of the electrode.…”
Section: Sem View Of Nanochannel Arraymentioning
confidence: 99%
“…The stability and consistency of the batch-microfabricated miniaturized Ag/AgCl RE were compared with previously reported Ag/AgCl REs 17,18,23,26 and were given in Table 2.…”
Section: Application Of Re In the Determination Of Nutrients In Watermentioning
confidence: 99%
“…The lifetime of an electrode is still a severe problem; therefore, studies on the salt bridge of REs have continued. 22,23 Pedrotti et al designed a high-density microporous polyethylene for a restricted ow junction between the internal and external electrolytes. The more favorable characteristics of a microporous membrane for the fabrication of miniaturized junctions are evident in a commercial Vycor membrane.…”
Section: Introductionmentioning
confidence: 99%
“…When the hole diameter was 20 mm, the potential change was 0.01 mV per day during the testing for two months. 23 In addition, some special applications and integration studies on Ag/AgCl REs have been reported. [24][25][26] Hayashi et al invented a RE to use in a ow-type measuring apparatus by incorporating an immobilized enzyme WE.…”
In this paper, a miniaturized Ag/AgCl reference electrode, which can be mass-produced by micro–nano manufacturing technology, was developed and demonstrated to have good stability and a long lifetime.
“…Compared with a traditional glass tube RE, the nanochannel array of the novel RE can signicantly decrease the loss of chloride ions in the KCl solution and effectively increase the service life of the RE. This shows that the dri rate of a RE reference potential is proportional to the diameter of the ion exchange channel, as obtained by Zevenbergen et al 23 The Ag/AgCl electrode with a thickness of about 0.5 mm was fabricated on a Pyrex 7740 glass using a silver sputtering li-off process and electrochemical treatment using a HCl solution. A hole with a 1.4 mm diameter was prepared using a laser for injecting a saturated KCl solution.…”
Section: Miniaturized Ag/agcl Re Designsupporting
confidence: 68%
“…A typical glass tube RE uses a semipermeable membrane to exchange the internal solution with the external water environment. Previous studies have used high-density microporous polyethylene 22 with a single micron hole 23 instead of the semipermeable membrane of the traditional glass tube Ag/AgCl electrode. This study gradually improved the performance of the electrode, especially the life of the electrode.…”
Section: Sem View Of Nanochannel Arraymentioning
confidence: 99%
“…The stability and consistency of the batch-microfabricated miniaturized Ag/AgCl RE were compared with previously reported Ag/AgCl REs 17,18,23,26 and were given in Table 2.…”
Section: Application Of Re In the Determination Of Nutrients In Watermentioning
confidence: 99%
“…The lifetime of an electrode is still a severe problem; therefore, studies on the salt bridge of REs have continued. 22,23 Pedrotti et al designed a high-density microporous polyethylene for a restricted ow junction between the internal and external electrolytes. The more favorable characteristics of a microporous membrane for the fabrication of miniaturized junctions are evident in a commercial Vycor membrane.…”
Section: Introductionmentioning
confidence: 99%
“…When the hole diameter was 20 mm, the potential change was 0.01 mV per day during the testing for two months. 23 In addition, some special applications and integration studies on Ag/AgCl REs have been reported. [24][25][26] Hayashi et al invented a RE to use in a ow-type measuring apparatus by incorporating an immobilized enzyme WE.…”
In this paper, a miniaturized Ag/AgCl reference electrode, which can be mass-produced by micro–nano manufacturing technology, was developed and demonstrated to have good stability and a long lifetime.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.