“…6). DONALDSON (1983) found that Cooper cable failed by wire fracture if the pulleys were 5 mm radius, as here, or by rubber failure, if the radius was 3 ram. The failure of cable 2 does not confound that result because this cable was subjected to dry running, which causes chafing where the cable rubs against itself at the top of the pulley.…”
Section: Noise Due To the Cablementioning
confidence: 90%
See 2 more Smart Citations
“…6). DONALDSON (1983) found that Cooper cable failed by wire fracture if the pulleys were 5 mm radius, as here, or by rubber failure, if the radius was 3 ram. The failure of cable 2 does not confound that result because this cable was subjected to dry running, which causes chafing where the cable rubs against itself at the top of the pulley.…”
Section: Noise Due To the Cablementioning
confidence: 90%
“…A cable testing machine was built, similar to that used originally (DONALDSON, 1983). Five cables were clamped at both ends in a tank of isotonic saline at room temperature.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…To implant IMDs in human body without complications, it has several critical requirements such as autoclavability, nontoxicity, biocompatibility and flexibility to shape freely to the change of patient's posture [1]. Furthermore, the IMDs should be safe, robust, and long-term operable without failure.…”
Section: Introductionmentioning
confidence: 99%
“…This is because available implantable multi-core cables were not intended to carry signals of this bandwidth and in this band they have significant resistance, inductance and capacitance. In this project a five core Cooper cable [19] was used, having a diameter of 2 mm and consisting of five helically wound cores of single strand platinum-iridium wire: this type of cable were designed to be small in diameter, very flexible and immune to fatigue (see photograph in Fig. 5).…”
Section: Introductionmentioning