Using isopropyl alcohol impregnated disinfection caps in the neonatal intensive care unit can cause isopropyl alcohol toxicity

Abstract: The SwabCap altered the valves' appearance and allowed significant amounts of isopropyl alcohol to be injected. It should not be used for neonates without further research.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

2
18
0
2

Year Published

2016
2016
2023
2023

Publication Types

Select...
8
1

Relationship

0
9

Authors

Journals

citations
Cited by 13 publications
(22 citation statements)
references
References 13 publications
2
18
0
2
Order By: Relevance
“…The main result of this in vitro study is that we can confirm the results from Sauron et al that there is a risk of IPA leakage when IPA caps are used [ 9 ]. This is troublesome as alcohol caps have been shown to be effective and are considered to be an important tool against catheter-associated infections [ 5 8 ].…”
Section: Discussionsupporting
confidence: 81%
See 1 more Smart Citation
“…The main result of this in vitro study is that we can confirm the results from Sauron et al that there is a risk of IPA leakage when IPA caps are used [ 9 ]. This is troublesome as alcohol caps have been shown to be effective and are considered to be an important tool against catheter-associated infections [ 5 8 ].…”
Section: Discussionsupporting
confidence: 81%
“…Based on the previously mentioned study [ 9 ] we decided to test two commercially available alcohol caps together with two hubs commonly used in neonatal intensive care in order to evaluate potential leakage from alcohol caps. For comparison we also evaluated the leakage of alcohol from the scrub-the-hub method.…”
Section: Introductionmentioning
confidence: 99%
“…Routine use of disinfection caps among adult high‐risk hematology oncology patients proved a reduction in both hospital‐acquired CLABSI and blood culture contamination, 34% and 63%, respectively 62 . Sauron et al 63 caution of its use in the neonatal population as laboratory testing of SwabCap (Excelsior Medical, LLC, Neptune, NJ), 1 brand of IPA‐impregnated disinfection cap, after a 96‐hour contact with SmartSite or CareSite valves, compromised the valves' appearance. Although toxic plasmatic levels are unknown, based on calculated assumptions of tolerated levels, the IPA levels were almost doubled.…”
Section: Prevention Strategies—disinfection Capsmentioning
confidence: 99%
“…Auf die Gefahr des Bakterieneintrags durch die Membranen der NFC bei ausbleibender Desinfektion vor Verwendung wird explizit hingewiesen. Von Isopropanol-freisetzenden Verschlusskappen auf NFCs wird abgeraten, weil sie die Integrität von NFCs beeinträchtigen können und es bei Injektionen über das NFC zu einem Eintrag von Isopropylalkohol in die Blutbahn kommen kann [15]. Erwähnt wurde außerdem eine Studie, die einen signifikanten Rückgang der CABSI Rate nach Einführung eines "komplexen geschlossenen Medikationssystems", bei dem unter anderem Spritzen bis zum Wechsel am System verbleiben, zeigte [1].…”
Section: Nadelfreie Konnektionsventile (Nfc) Und Konnektionsstellenunclassified