Using multileaf collimator interleaf leakage to extract absolute spatial information from electronic portal imaging device images

Abstract: Electronic portal imaging devices (EPIDs) are potentially valuable tools for linear accelerator quality assurance and for measuring and analyzing geometric variations in radiation treatment delivery. Geometric analysis is more robust if referenced against an absolute position such as the isocenter (collimator axis of rotation), allowing the observer to discriminate between various setup errors and jaw or multileaf collimator (MLC) calibration errors. Unfortunately, mechanical instabilities in EPIDs make such a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2012
2012
2021
2021

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(2 citation statements)
references
References 17 publications
0
2
0
Order By: Relevance
“…If the lines were not vertical, the skewness of the system was determined from the slope of the lines (α). To differentiate between the leaf bank skewness relative to the collimator and the effects of collimator angle misalignment and EPID rotation in its plane, the Radon transform 47 was applied to one of the subareas between the pickets [e.g., A 3 in Fig. 1(c)] to determine the orientation of interleaf leakages (β).…”
Section: Iib3 Data Processingmentioning
confidence: 99%
“…If the lines were not vertical, the skewness of the system was determined from the slope of the lines (α). To differentiate between the leaf bank skewness relative to the collimator and the effects of collimator angle misalignment and EPID rotation in its plane, the Radon transform 47 was applied to one of the subareas between the pickets [e.g., A 3 in Fig. 1(c)] to determine the orientation of interleaf leakages (β).…”
Section: Iib3 Data Processingmentioning
confidence: 99%
“…In a previous study, Samant et al (2002) measured the shift in MLC leaf banks relative to the central axis using EPID images at zero gantry angle assuming their camera-based EPID had a rigid geometry (Samant et al 2002). In practice, the EPID cannot be considered as a mechanically stable structure during gantry rotations in arc treatments (Gao et al 2007, Ansbacher 2006, Rowshanfarzad et al 2012. In addition, the detection precision was restricted to the size of EPID pixels (Yang and Xing 2004).…”
Section: Introductionmentioning
confidence: 99%